A self-contained, compact, vertical or horizontal unidirectional flow module. The Envirco VLF 797/HLF 797 is an economical, multipurpose, unidirectional flow module that is completely self-contained. It provides an ISO Class 5 (Class 100) environment within the work area that complies with the latest IEST Recommended Practices. Assembly required.

Horizontal or Vertical Airflow
Using an existing work surface, the unique design of the VLF 797/HLF 797 enables quick and easy assembly to either a horizontal or vertical laminar airflow configuration. The positive pressure laminar airflow provides product protection only. Air from within the cabinetmore
Unit of Measure

Specifications

BTU/Hr

1350

Airflow

90 feet per minute ± 10 FPM (0.45 meters per second) average velocity measured 6 inches (152.4mm) from the diffuser screen. Uniformity ±20 % of average or better.

Power Requirements

4.0 A

Shipping Weight

150 lbs80 kg

Construction

Aluminum cabinet with front panel finished in corrosion resistant polyurethane enamel.

Filters

Aluminum framed High Efficiency Particulate Air (HEPA) type, with a minimum efficiency of 99.99% at 0.3 micron, aerosol challenged.

Prefilter

Washable type, polyurethane foam - 30% ASHRAE.

Motor/Blower Assembly

Direct drive, continuous duty 1/5 Hp motor, with sealed-for-life bearing and overload protection. Motor/blower assembly is electronically balanced (dynamically) and mounted on vibration dampening rubber grommets.

Speed Controller

Solid-state speed control for adjusting airflow velocity utilizing an easy to clean, membrane touchpad controller.
Electrical1 Standard 15 Amp, 115V, single phase, 60 Hz, power cord with 3-prong grounded plug. Separate ON/OFF switch on the top/back of the unit in addition to the touch pad controller. Fluorescent lighting fixture provides 45 ft. candles at the work surface.

Options

N/A Shroud for MAC 10 Unit Ultra-Violet Light Option - An optional UV light is available for surface decontamination. An over-sized polycarbonate front panel is utilized to prevent accidental exposure on UV light to personnel. The HLF configuration does not support the UV option.

Applications

N/A Critical Sample Preparation Hyperatlinetation I.V. Admixtures Media Preparation Microelectronic Assembly Microscopy Analysis Optical Assembly Pharmaceutical Compounding Sterility Testing Syringe Filling Tissue Culture

  • 1 Amps @ 115V/60 Hz